Nosso grupo organiza mais de 3.000 Séries de conferências Eventos todos os anos nos EUA, Europa e outros países. Ásia com o apoio de mais 1.000 Sociedades e publica mais de 700 Acesso aberto Periódicos que contém mais de 50.000 personalidades eminentes, cientistas de renome como membros do conselho editorial.
Periódicos de acesso aberto ganhando mais leitores e citações
700 periódicos e 15 milhões de leitores Cada periódico está obtendo mais de 25.000 leitores
Bhoj Raj Sharma
Purpose: The purpose of this study is to prospectively compare the image quality and radiation dose in two different 256- slice MDCT using Filter Back Projection and iDose4 (Hybrid iterative reconstruction) algorithm craniocervical CT angiography.
Materials and methods: Thirty patients were randomly assigned into two groups (Group A and B), each comprising 15 patients. All patients underwent craniocervical CTA examination. Group A (10 men, 5 women; mean age 57.13 ± 11.27; age range 41-73 years) and Group B (9 men, 6 women; mean age 58.80 ± 12.55; age range 41-76 years) received different amount of X-ray tube potential and tube current and underwent on two different version of reconstruction algorithm. Then two blinded radiologist analyzed image quality of craniocervical CTA independently. They performed the subjective image quality and objective image quality assessment. The Effective Radiation Dose (ED) was calculated using CT dose volume Index (CTDIvol.), Dose-Length Product (DLP) and conversion coefficient for chest (conversion factor k=0.014 mSv mGy-1cm-1). We used chi-square test, and Maan-Whitney U test for non-parametric independent t test for subjective evaluation of image quality and to assess inter-observer reproducibility of the image quality by the subjective method between two observers, we used the interclass correlation test.
Result: The quantitative image quality of iterative reconstruction algorithm was significant compared to FBP. The mean image noise in group B was 0.51 times lesser than group A (P-value <0.001).The mean image scores in both the groups by radiologist 1 and radiologist 2 was (3.65 ± 0.48 vs. 3.33 ± 0.63 respectively, p=0.002) and (3.73 ± 0.45 vs. 3.40 ± 0.69 respectively, p=0.002). The radiation dose was found to be reduced by 55% in group B compared to group A (P<0.0001) in craniocervical CTA.
Conclusion: Hybrid Iterative Reconstruction (HIR) Algorithm significantly improves the image quality in craniocervical CT, as well as helps to lower the radiation dose while maintaining the image quality.